Sunbeam Blood Pressure Monitor 7659 User Manual

INSTRUCTION MANUAL  
Automatic  
Digital Blood  
Pressure  
Monitor  
Model 7659  
Sunbeam-Oster Household Products  
Before operating  
P.N.99935  
this unit  
please read these instructions completely.  
 
D o Read instructions carefully before  
using  
M o n i t o r  
2. Do Rest arm on a flat surface so that upper arm is at heart level when  
taking measurements.  
3. Do Remain calm, quiet and still when measurements are being taken.  
4. DO make sure that the cuff is in the proper position according to your  
instruction manual.  
5. DO Insure that the exhaust velocity is 2-5 mmHg per second at the systolic  
point each time a measurement is taken.  
6. DO Wait at least 10 minutes between taking measurements.  
7. DO Pump unit pressure 30 to 60 mmHg above normal systolic before  
beginning measurement.  
8. DO Consult your physician before adjusting your medication based on  
readings from this Monitor.  
9. DO Use fresh alkaline batteries.  
IO. DO Use a dry, soft cloth to clean the instrument.  
DO NOT move arms or hands when measurement is being taken.  
2. DO NOT talk or chew when measurement is being taken.  
DO NOT take a series of measurements without waiting at least 10 minutes  
between measurements.  
3.  
DO NOT adjust medication based on readings from this Monitor.  
4 .  
5. DO NOT make any adjustments to unit, there are no user serviceable parts.  
Call l-800-597-5978 for further instructions.  
6. DO NOT smoke during or at least 30 minutes prior to taking blood pressure  
measurements.  
7 .  
DO NOT drop or otherwise mechanically shock the Monitor. This is a delicate  
precision device.  
DO NOT use thinner, alcohol, benzene or wet cloth to clean unit.  
DO NOT subject to extremes in temperature, humidity, direct sunlight or dust.  
DO NOT fold the Arm Cuff or Rubber Cuff Hose tightly during storage, as  
such treatment may shorten the life of the components.  
8 .  
9 .  
10  
Manufactured to meet specifications set by the Association for the  
Advancement of Medical Instrumentation (AAMI)  
 
Wh a t  
is Bl ood Pr e s s ur e ?  
Blood pressure is a measurement of the force of blood flowing against the walls of the  
arteries. Arterial blood pressure is constantly changing during the course of the cardiac  
cycle. The highest pressure in the cycle is called the SYSTOLIC BLOOD PRESSURE;  
the lowest is the DIASTOLIC BLOOD PRESSURE. Both pressure readings, the SYS-  
TOLIC and DIASTOLIC, are necessary to enable a physician to evaluate the status of a  
patient’s blood pressure. Many factors such as physical activity, anxiety, or the time of day,  
can influence your blood pressure. Blood pressure is typically low in the morning and  
increases from afternoon to evening. It is lower in the summer and higher in the winter.  
Why Is It a Good Thing to Measure Blood Pressure at Home?  
Having one’s blood pressure measured by a doctor in  
a
hospital or a clinic, and group  
health checks, tend to stimulate nervousness in the subject and may even create high blood  
pressure. Also blood pressure varies in accordance with a variety of conditions,, and so  
judgement is not possible on the basis of a single measurement. (See Fig. 2)  
The blood pressure measured first thing in the morning after getting up, before taking any  
food, and with the subject still, is known as the fundamental blood pressure. In practice it  
is rather difficult to record the fundamental blood pressure, but to come as near as possible  
to measuring. the blood pressure in aan environment that is close to this, is why it is useful  
to take the measurement at home.  
. .  
Systolic:  
The highest point of blood pressure; level of blood pressure produced  
as your heart beats.  
Diastolic: The lowest point of blood pressure; level of blood pressure produced  
as your heart rests between beats.  
Millimeters of  
Blood pressure is measured in terms of the  
height of mercury in a column. Measurements are written one over the other.  
For example, a systolic presssure of 120  
is expressed as “120 over  
and diastolic pressure of 80  
IMPORTANT: Only a physician is qualified to interpret your blood pressure  
measurements, and no device can replace regular medical examinations by  
your physician. It is recommended that your physician raview your procedure  
for using this blood pressure monitor. Your physician should verify blood  
pressure measurements before making adjustments to medication.  
 
Standards for assessment of high or low blood  
have been established by the World Health  
chart.  
without regard to age,  
(WHO), as shown in this  
(Fig. 1)  
Hpryepsesrut ernes e type  
Normal blood  
pressure type  
1
4
0
Reference Material: Investigation into Adult Diseases Report by the Ministry of Health  
and Social Security, 1971.  
Individual blood pressures vary greatly both on a daily and a seasonal basis.  
These variations are even more pronounced in hypertense patients. Normally the  
blood pressure rises while at work and is at its lowest during the sleeping  
period. The graph below illustrates the variations in blood pressure over a whole  
day with measurement taken every five minutes.  
(Fig. 2)  
A
B
140  
80  
eep  
12 1 3 1 4 1 5 1 6 1 7 1 8  
1 9 2021 2 2 23 24  
1
2
3
4
5
6
7
Time  
PM  
AM  
Shown is data tor measurements taken every 5 minutes. The thick line represents sleep.  
The rise in blood pressure at 4 PM (A in the graph) and 12 PM (B in the graph) corre-  
spond to an attack of pain and sexual intercourse. (Beven, Honour  
1969.)  
Sci.  
 
DIAGRAM  
A
Systolic blood pressure  
display section  
Diastolic blood pressure,  
puise display section  
Main frame  
Pressure Memory Star[  
Power  
Hose  
connector  
setting switch  
Hose Plug  
- C u f f hose  
 
Condition/Cause  
Corrective action  
Display mark  
Mark  
in the  
Measurement in progress  
remain quiet.  
surement condition and  
Measurement in  
flashes when pulse is  
detected.  
progress  
Mark flashes when power is Automatic exhaust is  
applied and there is air  
remaining in the cuff.  
formed.  
Exhaust  
-
-
Measurement has begun  
but the pressure was  
insufficient.  
Raise  
A
pressure 30-40  
more and measure once  
again.  
Insufficient  
pressure  
Appears when the battery  
voltage is excessively tow.  
Replace all four batteries  
with new ones.  
Replace  
batteries  
Appears when the blood  
pressure could not be  
obtained accurately.  
error  
Exhaust the air from the  
cuff, and re-apply the cuff  
properly and re-measure.  
(Wait  
re-measure)  
minutes before  
Appears when the pulse  
rate could not be obtained  
accurately.  
Pulse error  
 
. .  
Battery Installation  
Remove the battery cover and insert oatteries into the battery compartment as shown, taking  
extreme tare that the polarities  
+
and  
-
are observed. (Fig. 3)  
(Fig. 3)  
2. Attaching the Arm Cuff  
Wind the arm cuff around the left Upper arm tightly (for best results) SO that the bottom  
edge of the cuff is positioned one inch above the elbow joint.  
(Fig. 4)  
*Do not roll up a shirt or jacket.  
It may cause constriction of the Upper  
arm and proper measurement results  
Will not be obtained. (Fig. 5)  
3. Measuring Posture  
Remain in a seated or reclining position.  
Do not move your arm while measuring.  
(Fig. 5)  
 
l
l
When power is turned  
When the "o" is displayed the unit is ready for measurement  
in the arm cuff by squeezing it or removing the hose plug.  
ON,  
all display marks appear for approximately 1 second.  
mark appears, release air  
Approx. 1 second  
2. Attach the arm cuff to the left arm for best results.  
3. Pressurize the arm cuff with the inflation bulb.  
l
Pressurize the arm cuff to approximately 30-40  
above the estimated systolic blood  
pressure to be measured.  
I
I
4. When Pressurization is Completed, the automatic exhaust mechanism will  
gradually reduce arm cuff pressure, and the  
mark will appear indicating that  
measurement is in progress.  
1) Pressure value appears on the  
and exhaust speed on the right side of the display.  
*Remain still during measurement.  
2) When a pulse is detected, the pressure value moves to the right side of the display. At  
this time the  
mark flashes at the same rate as pulse rate, and the buzzer  
sounds when the pulsation is sufficient.  
F l a s h i n g
3) When the measurement is completed and buzzer sounds (beep), Systolic pressure is  
displayed on the left and Diastolic pressure on the right side of the display.  
2-3 seconds later, pressure value and pulse appear alternately on the display.  
*When a measurement is made with insufficient pressure the “Err” mark appears. Pressurize to a  
valve 30-40  
higher than that previously used and perform the measurement again.  
5. Push the air release valve to completely release the air from within the cuff.  
6. Turn Power OFF  
Press the Power switch to turn power OFF. (See Diagram A)  
7. Previous Measurement Display  
When the Memory switch is pressed, the previous measurement value is shown for  
about 5 seconds.  
8. Automatic Power OFF Function  
If this unit is left on after measurement an Automatic Power-OFF function turns the  
power OFF after approximately 2 min. 30 seconds. To continue measurement press  
the power switch turning the unit ON.  
 
Name and model number  
system  
Pressure Meter 7659  
Measuring method  
Power source  
Oscillometric method  
Type AA batteries x 4  
mmHg (Pressure)  
Measuring range  
(Pulse)  
-3 mmH or 2% whichever greater (Pressure)  
-5% (Pu se)  
Pressurization  
Manual pressurization  
Constant-Air-Release Valve System  
Manual air release valve  
Deflation  
Memory  
Built-in memory enabling display of the previously measured value and  
currents measurements.  
Automatic Power OFF  
To be automatically cut off after 2 minutes 30 seconds of nonuse to  
save energy.  
Battery life  
Approx. 12 months with  
3
min. usage per day  
Operating environment  
Storage environment  
Dimensions  
50” to  
to  
less than 85% RH  
less than 85% RH  
3”(D) x  
10.5 oz (including 4 batteries)  
Weight  
.
 
MODEL 7659  
REPLACEMENT PARTS  
AND ACCESSORIES  
Description  
Price  
47834-042  
47834-033  
47834-012  
47834-036  
47834-044  
47834-037  
9 9 9 3 5  
Regular Adult Cuff  
B a t t e r v C o v e r  
Replacement Hose  
Exhaust Valve Assembly  
Replacement Bulb  
Storage Case  
Instruction Book  
Blood Pressure Record  
Sheet (25/pack)  
1 2 . 9 5  
2
0
0
4 . 9 5  
4 . 9 5  
2 . 9 5  
4.50  
1 .oo  
3.95  
47834-006  
,
.
47834-038  
T
Connector  
4.95  
O r d e r  
1
Above prices include shipping and handling.  
Please print your name and address below.  
Street Address  
C i t y  
State  
Z i p  
Send to: Sunbeam-Oster Household Products  
117 Central Industrial Row Purvis, Mississippi 39475  
or call l-800-394-2326  
I
For Mastercard or VISA: ‘Circle one  
Mastercard  
VISA  
C a r d  
#
E x p . D a t e  
Authorization Signature  
 
LI MI TED ON YEAR WARRANTY  
This product is warranted for one year from the date of purchase against defects in  
material and workmanship. If the product fails to operate during this period, return  
it, prepaid, to Appliance Service Station, 117 Central Industrial Row,  
Purvis, MS 39475 for repair or replacement without charge at the manu-  
facturers option or call 1-800-597-5978. This warranty gives you specific legal  
rights, and you may also have other rights which vary from state to state.  
Warrantor: Sunbeam-Oster Household Products.  
@Sunbeam TM Health at Home 01992 Sunbeam-Oster Household Products  
Distributed by Sunbeam-Oster Household Products, Schaumburg, IL 60173  
MODEL7659  
REPLACEMENTPARTS  
AND ACCESSORI ES  
o me  
Please print your name and address below.  
City  
State  
Zip  
Include check or money order made payable to Sunbeam-Oster Household Products  
and mail to:  
Sunbeam-Oster Household Products  
117 Central Industrial Row Purvis, Missrssippi 39475  
 
MODEL 7659  
REPLACEMENT PARTS  
AND ACCESSORIES  
Description  
Price  
47834-042  
47834-033  
47834-012  
47834-036  
47834-044  
47834-037  
9 9 9 3 5  
Regular Adult Cuff  
B a t t e r v C o v e r  
Replacement Hose  
Exhaust Valve Assembly  
Replacement Bulb  
Storage Case  
Instruction Book  
Blood Pressure Record  
Sheet (25/pack)  
1 2 . 9 5  
2
0
0
4 . 9 5  
4 . 9 5  
2 . 9 5  
4.50  
1 .oo  
3.95  
47834-006  
,
.
47834-038  
T
Connector  
4.95  
O r d e r  
1
Above prices include shipping and handling.  
Please print your name and address below.  
Street Address  
C i t y  
State  
Z i p  
Send to: Sunbeam-Oster Household Products  
117 Central Industrial Row  
Purvis, Mississippi 39475  
I
or call l-800-394-2326  
For Mastercard or VISA: ‘Circle one  
Mastercard  
VISA  
C a r d  
#
E x p . D a t e  
Authorization Signature  
 

Sony Camcorder VCLHG1730A User Manual
Sony CRT Television KV PF21M70 User Manual
Sony Film Camera XC EU50CE User Manual
Sony Portable DVD Player MV 65ST User Manual
Sony Stereo System STRDH130 User Manual
Speco Technologies Computer Monitor VM 2105C User Manual
Sterling Dehumidifier stt 1600 User Manual
Sunbeam Charcoal Grill NBC1401 User Manual
SUPER MICRO Computer TV Video Accessories sc827 User Manual
Symbol Technologies Scanner LS2100 User Manual